![]() |
We refer to the transmit-receive response
as the
pulse-echo response. For the two-source geometry shown in
Figure 3.1, the pulse-echo response is contrasted
with the transmit-only response in Figure 3.2. The
corresponding k-space representations are shown in
Figure 3.3.
![]() |
![]() |
These solutions for a simple two-source geometry have been for the monochromatic case, i.e. for constant insonification with a single frequency. In these cases the far-field interference pattern in the observation plane technically stretches to infinity. If we consider a broadband case in which each source launches a short pulse, interference occurs only over a limited region in the observation plane, as shown in Figure 3.4. Outside of this region, the path length difference between the two sources and the observation point exceeds the pulse length, and hence interference no longer takes place.
![]() |
As shown in Figure 3.5, a two-element system with two-frequency output is sensitive to six spatial frequencies, with the redundant frequencies having twice the amplitude of the outliers. As the target gets further away, the outlying spatial frequencies move in as the point spread function spreads out.
![]() |
By extension, this example demonstrates why an aperture with some lateral extent is required to provide lateral resolution (i.e. lateral extent provides sensitivity to lateral spatial frequencies), and that given a finite target range and a very short pulse, the lateral resolution will only equal the axial resolution if the aperture becomes infinitely large.
![]() |
![]() |
With a large aperture and negligible kerf, the k-space magnitude response becomes smooth and continuous over a finite region. For a rectangular aperture producing a pulse with a Gaussian frequency response, this k-space frequency response is shaped like an asymmetrical mountain, with a tear-drop shaped footprint.
![]() |
The lateral pulse-echo transfer function for a rectangular aperture is
a triangle. The lateral k-space response is a sinc
function. For a broadband transducer, the axial response is often
approximated as a Gaussian centered at the transducer's center
frequency. These responses are shown in Figure 3.10. The
lateral pulse-echo beam pattern and k-space response is for
this aperture are seen to be a product in space and a convolution in
frequency of the transmit-only cases, respectively, as shown in Figure
3.11.
The axial response is an acoustic pulse that becomes progressively
stretched off axis, with interference increasing progressively off
axis.
![]() |
![]() |
The triangular lateral response is due to the presence of a larger number of elements at close spacing, while a smaller number of elements are placed at large spacing. Hence the weighting of the frequency response reflects the weighting of the sampling of spatial frequencies in the aperture, and is directly related to spatial resolution, as shown in Figure 3.12. The elements at the ends of the array are creating the highest spatial-frequency signals. If you remove the elements from the ends, the spatial frequency response gets narrower. The limited spatial resolution that results is not a sampling effect, but the result of the filtering effect of propagation from a limited aperture. Hence this limit is referred to as the diffraction-limited resolution of the system. In a typical system with an f-number on the order of 2, the lateral to axial spatial frequency mismatch can be on the order of 10!
![]() |
For example, if the k-space regions of support of two echo patterns overlap, the patterns themselves will be highly correlated. On the other hand, if they have no overlap, the corresponding speckle patterns will be completely uncorrelated. This is not an approximation, but the equivalent of the correlation operation in two dimensions. It is rigorous within the constraints of the approximation underlying the k-space representation.
Given the sinc
-shaped response of a rectangular aperture, we
observe a trade-off between lateral resolution and contrast.
Commercial machines apodize the aperture to reduce the sampling of
high lateral frequency components in the aperture. This reduces
spatial resolution but improves contrast by reducing the energy in the
side lobes of the sinc
.
One could attempt to apodize the aperture to increase the sampling of high lateral frequency components in the aperture. This would produce a sharper main lobe, but this improvement in lateral resolution would be offset by a degradation in contrast due to the emphasis of the side lobes. This approach is sometimes used in commercial scanners in a zoom mode.
Another trade-off is lateral resolution versus the depth-of-field
(DOF). Given
, DOF =
. Note that
this dependence is with the square of the aperture size. As shown in
Figure 3.13, the smaller the DOF, the smaller the
``sweet-spot'' of the transmit response. The DOF can be kept constant
on receive by using dynamic aperture growth on receive. On transmit,
this can only be achieved by using multiple transmit foci. This
reduces the frame rate by a factor proportional to the number of
transmit foci.
![]() |
The k-space examples shown up to this point have been flat phase, but under real conditions there is often has some phase profile across the aperture. This profile may be artificial, e.g. to achieve focusing and steering of the beam, or it may be a natural phase aberration arising from structures in the medium. Ideal focusing achieves perfect superposition of waves at the focus. In a real system, the focus phase profile is quantized. This produces phase quantization error at the focus, as shown in Figure 3.14. The magnitude of this error is determined by the transducer geometry and the scanner hardware.
![]() |
While focusing in the lateral dimension of a 1-D array can be applied electronically, focusing in the elevation dimension is typically achieved with an acoustic lens, which by its nature has a fixed focal length. Acoustic lenses are often quite severe, therefore the depth of field is poor in elevation. As a consequence the elevation dimension spatial resolution can vary by a factor of 10 within a typical range of interest. Clinicians are often not aware of this slice thickness problem.
Annular arrays are well focused in both dimensions, but can only used with mechanical scanners. These types of scanners cannot perform color-flow or simultaneous Doppler/B-mode imaging.
Target structures that scatter with spatial frequencies higher than
the extent of the system response are sub-resolution and are not
imaged by the system. The k-space representations of target
functions are related to their information content. Spatial
frequencies in the target function do not represent plate reflectors,
but rather are sinusoidal variations in scatterer reflectivity. For
example, the glomeruli in the kidney create a
cycles/mm
pattern. A schematic of this type of scattering structure is shown in
Figure 3.15. Resolving this structure requires
spatial resolution of at least 1/6 mm in the dimension along which the
structure's periodicity is oriented.
![]() |
In k-space this target is represented with an impulse-pair indicating the periodicity of the amplitude variation, as shown in Figure 3.16. In order to see this target with a minimal system, you have to transmit at the same spatial frequency.
![]() |
With a repeated structure, a comb function is produced, as shown in Figure 3.17. If this structure were rotated, one would need to launch the pulse along a similarly rotated axis in order to see the structure.
![]() |
Given a dominant axial scatterer spacing of
, consider the
predicted response of the system, shown schematically in Figure
3.18. Observe that this type of target produces
lobes in the echo frequency response. Note that the transmit-receive
process introduces a factor of two in the observed axial periodicity,
e.g. a dominant scatterer spacing of
in space is observed as a
spacing in transmit-receive
-space. Periodicity in the
scattering function produces periodicity in the echo spectrum. This
is the basis of some tissue characterization techniques, e.g. if
pathology is correlated with tissue structure at scales to which the
imaging system is sensitive, the associated echo spectra may carry
useful diagnostic
information[14,15,16,17,18,19,20].
![]() |
![]() |
The echogenicity of tissue also has a frequency dependence that can
run counter to frequency dependent attenuation. A Rayleigh scatterer
is a scatterer of size
. This echo intensity from this
type of scatterer has an
dependence. This relationship helps the
imaging process. For example, in the 1-5 MHz range, the echo from
blood is approximately 20 decibels below that from tissue, making
imaging and quantification of slow flow difficult. However, in the
20-30 MHz range blood cells become Rayleigh scatterers, such that the
echo from blood is as bright as from the tissue.
Finally, the target-array geometry can produce spatial-frequency weighting that is a function of position. More widely-spaced elements see shorter path length differences among scatterers along the beam axis, as shown in Figure 3.20. Therefore scatterers are effectively closer together. Therefore one needs a higher frequency to produce the same axial interference pattern as that observed with elements at the center of the array.
![]() |
A real-world factor not included in our theoretical discussion is the anisotropy of tissue. K-space was in fact first applied in ultrasound to describe tissue anisotropy[1]. Skeletal muscle has strong anisotropy due to its structure, but this type of muscle is a little interest in the ultrasound imaging. Cardiac muscle, on the other hand, shows an anisotropy that changes with both of the location in the heart and over the cardiac cycle.
Miller et. al [21] have shown that the cyclic variation in backscatter due to this anisotropy is different in healthy versus pathologic cardiac tissue. This is considered a potential means of diagnosis, although it is difficult to apply this method in a controlled manner, as the orientation of the sound beam and of the organ itself affects the results.