  1. You are given a 1-D transducer with 16 elements that operates at 8 MHz.  The elements are lambda wide, 1 mm high, and the overall length of the array is 20 mm. The focal point is at 3 cm; the transducer fractional BW is 25%.
    1.      (a) What is the lateral and axial resolution at the focal point (-6dB FWHM)?
           (b) What is the position and level of the first sidelobe?
           (c) What is the position and level of the first grating lobe?
       
  2. The above transducer was sparsely populated.  We now redo the same transducer so that it has 64 elements, but all other parameters are the same.
    1. (a) What is the highest lateral spatial frequency in the K-space image (use a 95% cutoff)?  Based on this, what beam spacing should you use? (This is lateral increment, or sampling).
      (b) What is the % degradation of lateral and axial resolution 1 cm beyond the focus?

  3. Design an array that maintains a 500 micron resolution at the 3 cm focal point.  The available transducer BW is 50%, and the tissue attenuation is 1 dB/cm*MHz.  You  are limited to 32 elements.  Assume that you can distinguish blood from tissue (these differ by 30 dB).  Choose the center frequency so that the overall dynamic range is 100dB, and so that the blood/tissue distinction is maintained from the face of the array to the focal point.  Also, the design must provide grating lobes that are > 40 dB down.  Give all relevant specifications for your design, and include relevant plots to demonstrate that your design meets the requirements.